Improved precision for the autoanalyser determination of serum cholesterol
Modifications are described for the Technicon N 24A method 1 for the analysis of serum cholesterol by the AutoAnalyser. The precision is thereby improved so that over a period of about 6 months the coefficient of variation was 1.75% (control serum concentration 200 mg/100 ml, 1 SD = 3.5). The use of...
Gespeichert in:
Veröffentlicht in: | Clinica chimica acta 1970-10, Vol.30 (1), p.47-52 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Modifications are described for the Technicon N 24A method
1 for the analysis of serum cholesterol by the AutoAnalyser. The precision is thereby improved so that over a period of about 6 months the coefficient of variation was 1.75% (control serum concentration 200 mg/100 ml, 1 SD = 3.5). The use of preparative thin-layer chromatography allows for the separate determination of free and ester fractions when required. |
---|---|