Improved precision for the autoanalyser determination of serum cholesterol

Modifications are described for the Technicon N 24A method 1 for the analysis of serum cholesterol by the AutoAnalyser. The precision is thereby improved so that over a period of about 6 months the coefficient of variation was 1.75% (control serum concentration 200 mg/100 ml, 1 SD = 3.5). The use of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinica chimica acta 1970-10, Vol.30 (1), p.47-52
Hauptverfasser: Fosbrooke, Audrey S., Pringle, G.A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Modifications are described for the Technicon N 24A method 1 for the analysis of serum cholesterol by the AutoAnalyser. The precision is thereby improved so that over a period of about 6 months the coefficient of variation was 1.75% (control serum concentration 200 mg/100 ml, 1 SD = 3.5). The use of preparative thin-layer chromatography allows for the separate determination of free and ester fractions when required.
ISSN:0009-8981
1873-3492
DOI:10.1016/0009-8981(70)90191-9