Air Leaks as a Source of Distortion in Apexcardiography

Snudi air leaks have been observed to be frequently present in a popular crystal transducer and to cause significant distortion of the apexcardiogram. Such leaks introduce a short time constant thereby differentiating the wave form and producing: (1) peaked and amplified A waves; (2) early fall off...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chest 1970-02, Vol.57 (2), p.163-169
Hauptverfasser: Kastor, John A., Aronow, Saul, Nagle, Robert E., Garber, Thomas, Walker, Harriet
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Snudi air leaks have been observed to be frequently present in a popular crystal transducer and to cause significant distortion of the apexcardiogram. Such leaks introduce a short time constant thereby differentiating the wave form and producing: (1) peaked and amplified A waves; (2) early fall off of the systolic wave; (3) increase in slope and relative duration of the rapid filling wave. Leaking was found to develop at the seal between the transducer housing and a gasket and was easily eliminated with the application of silicone lubricant. A test “square wave” was produced by quickly clamping the tube connection from bell to transducer, and a short time constant thereby observed as a clue to the presence of leaking. The possibility of such leaks and of certain other physical limitations discussed in this paper support the need for a new device which can simultaneously reproduce the phonocardiogram and apexcardiogram.
ISSN:0012-3692
1931-3543
DOI:10.1378/chest.57.2.163