The Evaluation of an Ultrasonic Flow Detector for the Assessment of Peripheral Vascular Disease
The circulation in the lower limbs of 50 patients with vascular disease was assessed transcutaneously with a commercially available ultrasonic flow detector. In 66 limbs, the results were compared with arteriograms. The findings were entirely correct in 57 limbs, and in every limb a reliable indicat...
Gespeichert in:
Veröffentlicht in: | Cardiovascular research 1969-10, Vol.3 (4), p.503-509 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The circulation in the lower limbs of 50 patients with vascular disease was assessed transcutaneously with a commercially available ultrasonic flow detector. In 66 limbs, the results were compared with arteriograms. The findings were entirely correct in 57 limbs, and in every limb a reliable indication of the adequacy of the blood supply was obtained. |
---|---|
ISSN: | 0008-6363 1755-3245 |
DOI: | 10.1093/cvr/3.4.503 |