Scanning multiple slit assembly: a practical and efficient device to reduce scatter
A scanning multiple slit assembly consisting of an array of long narrow beam-defining slits coupled with scatter-eliminating slots beneath the patient is a feasible, practical, and efficient method of reducing scatter and increasing contrast and image quality in diagnostic radiology. Measurements of...
Gespeichert in:
Veröffentlicht in: | American journal of roentgenology (1976) 1977-09, Vol.129 (3), p.497-501 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A scanning multiple slit assembly consisting of an array of long narrow beam-defining slits coupled with scatter-eliminating slots beneath the patient is a feasible, practical, and efficient method of reducing scatter and increasing contrast and image quality in diagnostic radiology. Measurements of the ratio of scattered-to-primary radiation transmitted by a scanning multiple slit assembly and a conventional high ratio grid under similar conditions showed the relative intensity of scatter transmitted by the slit assembly to be only one-third that of the grid. A noticeable improvement in contrast was obtained using the slit assembly, without increase in patient exposure. The design, construction, and clinical implications of such a scatter-eliminating technique are discussed. |
---|---|
ISSN: | 0361-803X 1546-3141 |
DOI: | 10.2214/ajr.129.3.497 |