Scanning multiple slit assembly: a practical and efficient device to reduce scatter

A scanning multiple slit assembly consisting of an array of long narrow beam-defining slits coupled with scatter-eliminating slots beneath the patient is a feasible, practical, and efficient method of reducing scatter and increasing contrast and image quality in diagnostic radiology. Measurements of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of roentgenology (1976) 1977-09, Vol.129 (3), p.497-501
Hauptverfasser: Barnes, GT, Brezovich, IA, Witten, DM
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A scanning multiple slit assembly consisting of an array of long narrow beam-defining slits coupled with scatter-eliminating slots beneath the patient is a feasible, practical, and efficient method of reducing scatter and increasing contrast and image quality in diagnostic radiology. Measurements of the ratio of scattered-to-primary radiation transmitted by a scanning multiple slit assembly and a conventional high ratio grid under similar conditions showed the relative intensity of scatter transmitted by the slit assembly to be only one-third that of the grid. A noticeable improvement in contrast was obtained using the slit assembly, without increase in patient exposure. The design, construction, and clinical implications of such a scatter-eliminating technique are discussed.
ISSN:0361-803X
1546-3141
DOI:10.2214/ajr.129.3.497