Measurement of Atherosclerotic Luminal Irregularity and Obstruction By Radiographic Densitometry
This report describes a procedure to quantify stenosis and to detect enface plaques in single plane angiograms. We derived an optical density function proportional to local arterial chord length from arteriographs over a wide range of conditions. Standard radiographs were made of silicone rubber bol...
Gespeichert in:
Veröffentlicht in: | Investigative radiology 1977-07, Vol.12 (4), p.307-313 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This report describes a procedure to quantify stenosis and to detect enface plaques in single plane angiograms. We derived an optical density function proportional to local arterial chord length from arteriographs over a wide range of conditions. Standard radiographs were made of silicone rubber bolus used as the radiopaque medium in cadaver arteries. Replicate arterial casts filled with iodine-bearing contrast medium were radiographed for comparison. Identical atherosclerotic areas in both film sets were scanned with a microdensitometer and vessel chord lengths used to construct arterial cross sections for comparison with vessel casts. Cross-sectional areas derived from cadaver and phantom films were highly correlated with those derived from cast slices (r=0.997). Thus, luminal cross-sectional areas were accurately estimated independent of plaque location, and accuracy of chord length measurement was ± 0.28 mm in vessels four to 10 mm in diameter. |
---|---|
ISSN: | 0020-9996 1536-0210 |
DOI: | 10.1097/00004424-197707000-00002 |