The Humongotron--A Scintillation-Camera Transaxial Tomograph

An emission transaxial tomographic system using a scintillation camera as the detector is described. The unit allows accurate positioning of a scintillation camera's detector at any angle around a patient in order to obtain the multiple projection images needed for transaxial tomography, and it...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of nuclear medicine (1978) 1977-04, Vol.18 (4), p.381-387
Hauptverfasser: Keyes, John W., Jr, Orlandea, Nicolae, Heetderks, William J, Leonard, Patrick F, Rogers, W. L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An emission transaxial tomographic system using a scintillation camera as the detector is described. The unit allows accurate positioning of a scintillation camera's detector at any angle around a patient in order to obtain the multiple projection images needed for transaxial tomography, and it is capable of imaging any area of the body. The camera can also be used for all types of conventional imaging procedures. Image processing is performed by a small on-line computer. A convolution algorithm and a mathematical technique for approximate absorption correction are used to obtain high-resolution and high-contrast images with good quantitative accuracy. The operation of the system is described and representative phantom and patient studies are presented to illustrate the capabilities of the system.
ISSN:0161-5505
1535-5667