Positive correlation of optical density at 650 nm. with lecithin/sphingomyelin ratios in amniotic fluid
In this study, we have attempted to correlate optical density measurements of amniotic fluids with L/S ratios. We may conclude, with over a 98 per cent accuracy, that fluids having optical density readings of 0.15 and above, at 650 nm. will have an L/S ratio over 2.0. Fluids having optical density r...
Gespeichert in:
Veröffentlicht in: | American journal of obstetrics and gynecology 1978-04, Vol.130 (7), p.788-790 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this study, we have attempted to correlate optical density measurements of amniotic fluids with L/S ratios. We may conclude, with over a 98 per cent accuracy, that fluids having optical density readings of 0.15 and above, at 650 nm. will have an L/S ratio over 2.0. Fluids having optical density readings up to 0.05 will have L/S ratios of about 1.3. Finally, amniotic fluids having optical densities >0.05 and |
---|---|
ISSN: | 0002-9378 1097-6868 |
DOI: | 10.1016/0002-9378(78)90009-1 |