Cineangiographic measurements with the help of a vertebral grid system

Cineangiographic measurements of cardiac structures require knowledge of the magnification factor due to nonparallelity of the X‐ray beams. A new method is presented to determine the magnification factor from cineangiograms based on the patient's height and the height of three vertebral bodies...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Catheterization and cardiovascular diagnosis 1976, Vol.2 (4), p.353-359
Hauptverfasser: Wagner, H. R., Teske, D. W., Siu, T. Oswald
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Cineangiographic measurements of cardiac structures require knowledge of the magnification factor due to nonparallelity of the X‐ray beams. A new method is presented to determine the magnification factor from cineangiograms based on the patient's height and the height of three vertebral bodies and the adjacent intervertebral distances. The method enables measurement of cardiovascular structures from projected cineangiographic films where the appropriate magnification factor is missing. Furthermore, doubtful magnification factors can be checked out as to their accuracy.
ISSN:0098-6569
1097-0304
DOI:10.1002/ccd.1810020408