Improvement of accuracy in a high-capacity, six degree-of-freedom load cell : Application to robotic testing of musculoskeletal joints
This study investigated a previously unaccounted for source of error in a high-capacity, six degree-of-freedom load cell used in multi-degree-of-freedom robotic testing of musculoskeletal joints, an application requiring a load cell with high accuracy in addition to high load capacity. A method of c...
Gespeichert in:
Veröffentlicht in: | Annals of biomedical engineering 1999-11, Vol.27 (6), p.839-843 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study investigated a previously unaccounted for source of error in a high-capacity, six degree-of-freedom load cell used in multi-degree-of-freedom robotic testing of musculoskeletal joints, an application requiring a load cell with high accuracy in addition to high load capacity. A method of calibration is presented for reducing the error caused by changes in universal force-moment sensor (UFS) orientation within a gravitational field. Uncorrected, this error can exceed a magnitude of 1% of the full-scale load capacity-the manufacturer-stated accuracy of the UFS. Implementation of the calibration protocol reduced this error by approximately 75% for a variety of loading conditions. This improvement in load cell accuracy (while maintaining full load capacity) should improve both the measurement and control of specimen kinetics by robotic/UFS and other biomechanical testing systems. |
---|---|
ISSN: | 0090-6964 1573-9686 |
DOI: | 10.1114/1.236 |