Failure of Beal mitral valve prosthesis. Clinical and radiographic features

This report presents a case of severe dysfunction of an improved Model 104 Beall valve secondary to disc wear. This type of valve failure may be extremely difficult to detect by auscultation, echocardiography, or fluoroscopic examination. Prosthesis failure with resulting mitral insufficiency should...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of thoracic and cardiovascular surgery 1975-07, Vol.70 (1), p.163-165
Hauptverfasser: Jost, RG, McKnight, RC, Roper, CL
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This report presents a case of severe dysfunction of an improved Model 104 Beall valve secondary to disc wear. This type of valve failure may be extremely difficult to detect by auscultation, echocardiography, or fluoroscopic examination. Prosthesis failure with resulting mitral insufficiency should be considered in any patient with a Teflon disc Beall mitral valve in whom there is unexplained clinical deterioration.
ISSN:0022-5223
1097-685X
DOI:10.1016/s0022-5223(19)40395-4