A “depth-independent” collimator for use with 99mTc in both conventional and transverse section scanning

The design of a “depth independent” collimator for use with 99 m Tc is discussed. Since this is a novel design, not available commercially, experimental measurements of line spread function (L.S.F.), both in air and phantom material (Mix-D) and the modulation transfer function (M.T.F.) calculated fr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of nuclear medicine & biology 1975-07, Vol.2 (3), p.123-128
Hauptverfasser: Lakshmanan, A.V., Causer, D.A., Wilks, R.J., Taylor, C.G., Mallard, J.R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The design of a “depth independent” collimator for use with 99 m Tc is discussed. Since this is a novel design, not available commercially, experimental measurements of line spread function (L.S.F.), both in air and phantom material (Mix-D) and the modulation transfer function (M.T.F.) calculated from the L.S.F. are presented. The collimator is shown to have a good depth of focus and has been found to have an acceptable sensitivity for brain scanning. A pair of these collimators have been particularly useful for transverse section scanning of the brain.
ISSN:0047-0740
DOI:10.1016/0047-0740(75)90029-7