THE RESISTANCE OF ENTONOX VALVES

The resistances of thirteen Entonox valves were measured. Although there was a wide variation between them, this is unlikely to interfere with their performance when used in labour. During the passage of continuous high flows, freezing occurred in two valves, causing a reduction in flow after 5 minu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 1974-02, Vol.46 (2), p.145-148
Hauptverfasser: DAVIES, JUDITH M., HOGG, M.I.J., ROSEN, M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The resistances of thirteen Entonox valves were measured. Although there was a wide variation between them, this is unlikely to interfere with their performance when used in labour. During the passage of continuous high flows, freezing occurred in two valves, causing a reduction in flow after 5 minutes. It is very unlikely that this would occur in clinical use. Measurements of the resistances of eight of the valves were compared using a static tube or a wide-bore needle to measure pressure changes. There was no significant difference between the two methods at flows of up to 100 1./min.