THE RESISTANCE OF ENTONOX VALVES
The resistances of thirteen Entonox valves were measured. Although there was a wide variation between them, this is unlikely to interfere with their performance when used in labour. During the passage of continuous high flows, freezing occurred in two valves, causing a reduction in flow after 5 minu...
Gespeichert in:
Veröffentlicht in: | British journal of anaesthesia : BJA 1974-02, Vol.46 (2), p.145-148 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The resistances of thirteen Entonox valves were measured. Although there was a wide variation between them, this is unlikely to interfere with their performance when used in labour. During the passage of continuous high flows, freezing occurred in two valves, causing a reduction in flow after 5 minutes. It is very unlikely that this would occur in clinical use. Measurements of the resistances of eight of the valves were compared using a static tube or a wide-bore needle to measure pressure changes. There was no significant difference between the two methods at flows of up to 100 1./min. |
---|---|