Calculation of chromatographic enantioselectivity (α) is critically important in enantioselective chromatographic method development studies. The fact that α can only be calculated from isocratic elution conditions, whereas gradient elution conditions are predominantly used in method development scr...
Gespeichert in:
Veröffentlicht in: | Chirality (New York, N.Y.) N.Y.), 2011-02, Vol.23 (2), p.128-132 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Calculation of chromatographic enantioselectivity (α) is critically important in enantioselective chromatographic method development studies. The fact that α can only be calculated from isocratic elution conditions, whereas gradient elution conditions are predominantly used in method development screening, presents some problems for the use of α as a scoring indicator for automated, intelligent enantioselective chromatography method development systems. In this study, an empirical algorithm was developed to estimate α at isocratic conditions based upon information collected from a gradient elution. The algorithm was validated for SFC applications and has been shown to accurately predict enantioselectivity for a wide variety of racemic test analytes eluted on different chiral column and mobile phase conditions. Chirality, 2011. © 2010 Wiley‐Liss, Inc. |
---|---|
ISSN: | 0899-0042 1520-636X |
DOI: | 10.1002/chir.20887 |