A Device for Ensuring the Neochordae Replacement in Mitral Valve Repair

We have developed a new device for maintaining artificial chordae at the appropriate length during the tying of Gore-Tex sutures (W. L. Gore and Assoc, Flagstaff, AZ). This double-armed, double-hooked device is inserted through the loop formed by the neochordae, which is anchored in the papillary mu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 2010-12, Vol.90 (6), p.2071-2072
Hauptverfasser: Iida, Hiroshi, MD, PhD, Sunazawa, Toru, MD, Doi, Atsuo, MD, Ishida, Keiichi, MD, PhD, Irabu, Shinichiro, MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have developed a new device for maintaining artificial chordae at the appropriate length during the tying of Gore-Tex sutures (W. L. Gore and Assoc, Flagstaff, AZ). This double-armed, double-hooked device is inserted through the loop formed by the neochordae, which is anchored in the papillary muscle and passed through the prolapsing segment. The device pulls up both leaflets and maintains the neochordae at the same length as that of the opposing normal chordae. The prolapsed leaflet is suspended at the same height as the facing leaflet, enabling the accurate and reproducible placement of neochordae.
ISSN:0003-4975
1552-6259
DOI:10.1016/j.athoracsur.2009.12.085