Enzymatic assay for creatinine with fixed-time kinetics on a centrifugal analyzer

A method is described for enzymatic measurement of serum creatinine using a centrifugal analyzer. The method employs fixed-time kinetics and a true serum blank. The method required 100 microliters of serum and is linear to 90 mg per L. Day-to-day precision studies revealed a standard deviation (coef...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of clinical and laboratory science 1984-01, Vol.14 (4), p.298-303
Hauptverfasser: MARGREY, M, MARGREY, K, BRUNS, D. E, BOYD, J. C, FORTIER, G. A, RENOE, B. W, SAVORY, J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method is described for enzymatic measurement of serum creatinine using a centrifugal analyzer. The method employs fixed-time kinetics and a true serum blank. The method required 100 microliters of serum and is linear to 90 mg per L. Day-to-day precision studies revealed a standard deviation (coefficient of variation) of less than +/- 0.5 mg per L (7.4 percent) at a mean creatinine concentration of 6 mg per L and +/- 0.5 mg per L (1.8 percent) at a mean value of 28 mg per L. The method was free of interferences from acetoacetic acid, ascorbic acid, bilirubin, and hemoglobin. The speed, precision, and accuracy of the method suggest that it represents a useful alternative to methods using the alkaline picrate procedure.
ISSN:0091-7370
1550-8080