A simple technique to accurately determine intra-abdominal pressure

The determination of intra-abdominal pressure (IAP) may be useful in a variety of clinical situations. Despite this, invasive IAP monitoring is seldom performed due to the risks involved. Using a standard canine model of increased IAP, we evaluated the accuracy of transurethral bladder catheter pres...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Critical care medicine 1987-12, Vol.15 (12), p.1140-1142
Hauptverfasser: IBERTI, THOMAS J, KELLY, KATHY M, GENTILI, DAVID R, HIRSCH, STEVEN, BENJAMIN, ERNEST
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The determination of intra-abdominal pressure (IAP) may be useful in a variety of clinical situations. Despite this, invasive IAP monitoring is seldom performed due to the risks involved. Using a standard canine model of increased IAP, we evaluated the accuracy of transurethral bladder catheter pressure in reflecting IAP. Throughout the range of IAP studied (10 ± 5 to 70 ± 10 mm Hg), bladder pressure did not differ significantly from the direct measurement of IAP. We conclude that the measurement of bladder pressure using a standard transurethral bladder catheter provides an accurate determination of IAP.
ISSN:0090-3493
1530-0293
DOI:10.1097/00003246-198712000-00014