Venous Admixture to Radial Artery Blood Samples during Cardiopulmonary Bypass
The possibility that venous blood could be withdrawn through a radial artery cannula when venous pressure is high was examined in 30 patients during cardiopulmonary bypass for coronary artery bypass operation. Progressive desaturation of radial artery blood occurred when venous pressure was equal to...
Gespeichert in:
Veröffentlicht in: | Chest 1987-11, Vol.92 (5), p.836-838 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The possibility that venous blood could be withdrawn through a radial artery cannula when venous pressure is high was examined in 30 patients during cardiopulmonary bypass for coronary artery bypass operation. Progressive desaturation of radial artery blood occurred when venous pressure was equal to arterial pressure. Desaturation not only increased with duration of venous obstruction but also occurred immediately if the arterial tree was emptied after occlusion. We believe that these observations result from the transcapillary aspiration of deep venous blood and may produce a potential source of error in measuring arterial oxygen tension during cardiopulmonary resuscitation, cardiac tamponade, and profound right heart failure, as well as cardiopulmonary bypass. |
---|---|
ISSN: | 0012-3692 1931-3543 |
DOI: | 10.1378/chest.92.5.836 |