Atrial Fibrillation Induced by Testing Magnet Applied to VVI Pacemaker

The first known case of testing magnet‐induced atrial fibrillation in a patient with a VVI pacemaker is reported. A macro‐re‐entry mechanism is postulated, involving retrograde conduction from the ventricle through the AV node to the atrium following an asynchronous pacemaker stimulus.

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1984-03, Vol.7 (2), p.293-295
1. Verfasser: STALLER, BERNARD J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The first known case of testing magnet‐induced atrial fibrillation in a patient with a VVI pacemaker is reported. A macro‐re‐entry mechanism is postulated, involving retrograde conduction from the ventricle through the AV node to the atrium following an asynchronous pacemaker stimulus.
ISSN:0147-8389
1540-8159