Transmural surgical gown pressure measurements in the operating theater

Transmural gown pressures encountered when the surgeon comes into contact with a patient were measured in the operating theater. The surgical gown industry has assumed these pressures to be less than 5 psi in testing the efficacy of the gown and drape barrier material to impede bacterial transmissio...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of infection control 1991-06, Vol.19 (3), p.147-155
Hauptverfasser: Altman, K W, McElhaney, J H, Moylan, J A, Fitzpatrick, K T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Transmural gown pressures encountered when the surgeon comes into contact with a patient were measured in the operating theater. The surgical gown industry has assumed these pressures to be less than 5 psi in testing the efficacy of the gown and drape barrier material to impede bacterial transmission through its pores. In this study, pressure-sensitive contact film and resistive strain gauge recordings made from the surgeon's abdominal region and forearms indicated peak contact pressures in excess of 60 psi. This report indicates a need to reassess the basis of test utilization in evaluating barrier materials used in gowns and drapes.
ISSN:0196-6553
DOI:10.1016/0196-6553(91)90021-4