Physical and Biological Considerations Militating against the Use of Endoscopically Applied High-frequency Current in the Abdomen

Summary In general surgery, high-frequency current is a routine method of producing destructive heat for hemostasis. The so-called “skin effect” prohibits any electric reactions into the human nervous or muscle tissue. When using high-frequency current in the closed abdominal cavity, the nature of t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Endoscopy 1983-09, Vol.15 (5), p.283-288
1. Verfasser: Semm, K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Summary In general surgery, high-frequency current is a routine method of producing destructive heat for hemostasis. The so-called “skin effect” prohibits any electric reactions into the human nervous or muscle tissue. When using high-frequency current in the closed abdominal cavity, the nature of the current forces is to go unpredictable ways. This fact provokes unforeseen damage (bowel, ureter, skin burns) which are method-induced and cannot be excluded even with extreme care in the handling of the high-frequency current during tubal sterilization or other endoscopic haemostasis procedures. All biological and physical data are presented that support the demand that high-frequency current must be excluded from pelviscopy e.g. laparoscopy, and a new technique called “endocoagulation” is described, which can replace the intraabdominal use of high-frequency current during pelviscopy.