Physical and Biological Considerations Militating against the Use of Endoscopically Applied High-frequency Current in the Abdomen
Summary In general surgery, high-frequency current is a routine method of producing destructive heat for hemostasis. The so-called “skin effect” prohibits any electric reactions into the human nervous or muscle tissue. When using high-frequency current in the closed abdominal cavity, the nature of t...
Gespeichert in:
Veröffentlicht in: | Endoscopy 1983-09, Vol.15 (5), p.283-288 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Summary
In general surgery, high-frequency current is a routine method of producing destructive heat for hemostasis. The so-called “skin effect” prohibits any electric reactions into the human nervous or muscle tissue. When using high-frequency current in the closed abdominal cavity, the nature of the current forces is to go unpredictable ways. This fact provokes unforeseen damage (bowel, ureter, skin burns) which are method-induced and cannot be excluded even with extreme care in the handling of the high-frequency current during tubal sterilization or other endoscopic haemostasis procedures. All biological and physical data are presented that support the demand that high-frequency current must be excluded from pelviscopy e.g. laparoscopy, and a new technique called “endocoagulation” is described, which can replace the intraabdominal use of high-frequency current during pelviscopy. |
---|---|