A continuous-flow detector for cytochrome P-450 and cytochrome P-420
A spectrophotometric detector for automatically monitoring levels of cytochrome P-450 and cytochrome P-420 in chromatographic column effluents is described. Levels of cytochrome P-450 and P-420 as low as 30 pmol/ml can be detected above an absorbance background noise equivalent to 10 pmol/ml of thes...
Gespeichert in:
Veröffentlicht in: | Analytical biochemistry 1983-07, Vol.132 (1), p.26-33 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A spectrophotometric detector for automatically monitoring levels of cytochrome
P-450 and cytochrome
P-420 in chromatographic column effluents is described. Levels of cytochrome
P-450 and
P-420 as low as 30 pmol/ml can be detected above an absorbance background noise equivalent to 10 pmol/ml of these heme proteins. Spectra and the concentrations of cytochrome
P-450 and cytochrome
P-420 are determined every 5 sec. The automatic flow reactor and detector offer the following advantages over existing manual methods: (i) greatly reduced analysis time, (ii) measurement of a larger number of independent samples than is practical manually, (iii) simultaneous measurement of cytochrome
P-450 and its degradation product cytochrome
P-420 immediately after elution from the column, thus avoiding further sample degradation, and (iv) greatly increased accuracy and threshold resolution due to highly reproducible reaction conditions and constant optics. |
---|---|
ISSN: | 0003-2697 1096-0309 |
DOI: | 10.1016/0003-2697(83)90421-9 |