Multicentre Comparison of the Novo Biolabs Immunometric CK-MB Assay with Four Established Methods

The analytical performance of a new immunometric CK-MB assay from Novo Biolabs was evaluated in three hospitals. Comparisons were made with another immunometric assay, an immune inhibition/precipitation technique, an immune inhibition method, and a column separation method. The Novo assay exhibited...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of clinical biochemistry 1991-01, Vol.28 (1), p.73-77
Hauptverfasser: Wulkan, R W, van Loon, J, Sturk, A, Boerma, G J M, Verdegaal, M C J M, van den Berg, B, Souverijn, J H M, Sanders, G T B
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The analytical performance of a new immunometric CK-MB assay from Novo Biolabs was evaluated in three hospitals. Comparisons were made with another immunometric assay, an immune inhibition/precipitation technique, an immune inhibition method, and a column separation method. The Novo assay exhibited good precision (CV 3–7%) at all concentration levels tested (6–150 μg/L) and was found to be linear over the range 0–200 μg/L. Comparison of the Novo kit with the other methods using patients' sera showed acceptable correlation between the methods over a wide concentration range (r = 0·96–0·99). We conclude that the assay is a useful adjunct to the specific measurement of CK-MB, particularly in routine laboratories with a high throughput of such determinations.
ISSN:0004-5632
1758-1001
DOI:10.1177/000456329102800112