Multicentre Comparison of the Novo Biolabs Immunometric CK-MB Assay with Four Established Methods
The analytical performance of a new immunometric CK-MB assay from Novo Biolabs was evaluated in three hospitals. Comparisons were made with another immunometric assay, an immune inhibition/precipitation technique, an immune inhibition method, and a column separation method. The Novo assay exhibited...
Gespeichert in:
Veröffentlicht in: | Annals of clinical biochemistry 1991-01, Vol.28 (1), p.73-77 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The analytical performance of a new immunometric CK-MB assay from Novo Biolabs was evaluated in three hospitals. Comparisons were made with another immunometric assay, an immune inhibition/precipitation technique, an immune inhibition method, and a column separation method. The Novo assay exhibited good precision (CV 3–7%) at all concentration levels tested (6–150 μg/L) and was found to be linear over the range 0–200 μg/L. Comparison of the Novo kit with the other methods using patients' sera showed acceptable correlation between the methods over a wide concentration range (r = 0·96–0·99). We conclude that the assay is a useful adjunct to the specific measurement of CK-MB, particularly in routine laboratories with a high throughput of such determinations. |
---|---|