Clinical Evaluation of a Simple Demand Inhalation MDI Aerosol Delivery Device
Inhalation of medication is the preferred method for treating reversible airway obstruction; however, difficulties in the use of pressurized canisters frequently lead to suboptimal results. The Aerochamber (Monoghan Medical Corp) is a portable breath-actuated device that attaches to a mete red-dose...
Gespeichert in:
Veröffentlicht in: | Chest 1983-07, Vol.84 (1), p.36-41 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Inhalation of medication is the preferred method for treating reversible airway obstruction; however, difficulties in the use of pressurized canisters frequently lead to suboptimal results. The Aerochamber (Monoghan Medical Corp) is a portable breath-actuated device that attaches to a mete red-dose inhaler (MDI) and is designed to overcome many of the problems of aerosol delivery encountered by some patients. The attachment of this breath-actuated device to an MDI reduced pharyngeal deposition of aerosol 14-fold, but delivery of aerosol to intrapulmonary airways in normal subjects and patients with bronchitis remained unchanged. In a group of nine patients with stable asthma, inhalation of a bronchodilator aerosol using the breath-actuated device (Aerochamber) achieved effective bronchodilation similar to an optimally administered MDI. Advantages of the breath-actuated device (Aerochamber) include (1) aerosol delivery of medication whether or not the discharge of aerosol is synchronized with inhalation, (2) effective therapeutic response compared with optimally administered MDI; (3) greatly reduced deposition of aerosol in the upper airways, which might be expected to reduce adverse effects of steriods; and (4) universal application to all bronchodilator and steroid MDIs. |
---|---|
ISSN: | 0012-3692 1931-3543 |
DOI: | 10.1378/chest.84.1.36 |