In Vitro Videographic Comparison of Argon and Nd:YAG Laser Iridotomy
We used an in vitro technique with high-magnification video recording to evaluate from the posterior side of the iris the immediate sequence of events during argon and Nd:YAG laser peripheral iridotomy. The observed effects differed strikingly. The argon laser caused a gradual mounding up of iris pi...
Gespeichert in:
Veröffentlicht in: | American journal of ophthalmology 1991-05, Vol.111 (5), p.589-594 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We used an in vitro technique with high-magnification video recording to evaluate from the posterior side of the iris the immediate sequence of events during argon and Nd:YAG laser peripheral iridotomy. The observed effects differed strikingly. The argon laser caused a gradual mounding up of iris pigment epithelium with each successive energy application before final penetration. This effect was reduced but not eliminated with higher power levels. The Nd:YAG laser caused complete disruption and dispersal of the pigment epithelium with a single pulse of energy. Additionally, a multiple focal point configuration of the Nd:YAG laser was observed to produce a significantly larger iridotomy than a single focal point configuration for comparable energy settings. These observations may in part explain the observed clinical advantage of the Nd:YAG laser over the argon laser for creation of a patent iridotomy. |
---|---|
ISSN: | 0002-9394 1879-1891 |
DOI: | 10.1016/S0002-9394(14)73704-3 |