Intraoperative measurement of rotational stability of femoral components of total hip arthroplasty

High out-of-plane forces acting on the hip joint can produce important rotational micromotion of the femoral component. This micromotion at the prosthesis interface may be detrimental to the stability of the implant. In cementless femoral implants this could prevent bone ingrowth, and in the cemente...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical orthopaedics and related research 1991-05, Vol.266 (266), p.119-126
Hauptverfasser: HARRIS, W. H, MULROY, R. D, MALONEY, W. J, BURKE, D. W, CHANDLER, H. P, ZALENSKI, E. B
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:High out-of-plane forces acting on the hip joint can produce important rotational micromotion of the femoral component. This micromotion at the prosthesis interface may be detrimental to the stability of the implant. In cementless femoral implants this could prevent bone ingrowth, and in the cemented component this could cause generation of particulate debris, lysis, and loosening. The introduction of the torque wrench micrometer for assessment of intraoperative femoral component stability can quantify the initial stability of primary cementless femoral components and critically evaluate the stability (at either the initial or revision arthroplasty) of both cemented and cementless femoral components. It allows the surgeon to produce a known torque in the direction and magnitude of the out-of-plane forces that load the hip in vivo.
ISSN:0009-921X
1528-1132
DOI:10.1097/00003086-199105000-00019