We report preliminary experience with a newly designed chest tube (JCT), for evacuation of neonatal pneumothorax. The catheter has a unique pigtail configuration at the distal end, intended to simplify placement and minimize chest wall and lung trauma by reduced tube size and depth and insertion. Th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical pediatrics 1991-02, Vol.30 (2), p.85-87
Hauptverfasser: Jung, August L., Nelson, James, Jenkins, M. Bruce, Hodson, W. Alan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report preliminary experience with a newly designed chest tube (JCT), for evacuation of neonatal pneumothorax. The catheter has a unique pigtail configuration at the distal end, intended to simplify placement and minimize chest wall and lung trauma by reduced tube size and depth and insertion. Thirty-eight JCTs were placed in neonates with pneumothoraces. Neonates' birth weights ranged from 400 to 3,595 grams. All 38 tubes immediately relieved clinical signs of pneumothoraces. Thirty-five (92%) tubes immediately fully evacuated the pneumothoraces as evidences on chest radiograph. Twelve pneumothoraces partially reoccurred at a mean of 24 hours following JCT placement. These tubes were either irrigated or replaced. This newly configured chest tube functions effectively in the treatment of neonatal pneumothorax.