Comparison of the fast-neutron sensitivity of a Geiger-Muller counter using different techniques
The sensitivity of an energy-compensated Geiger-Müller counter to neutrons with approximate energies of 0.6, 5 and 15 MeV, relative to its sensitivity to gamma rays (kU-value), has been determined at different centres in Europe. The time-of-flight, angular difference, associated-particle coincidence...
Gespeichert in:
Veröffentlicht in: | Physics in medicine & biology 1982-01, Vol.27 (1), p.91-96 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The sensitivity of an energy-compensated Geiger-Müller counter to neutrons with approximate energies of 0.6, 5 and 15 MeV, relative to its sensitivity to gamma rays (kU-value), has been determined at different centres in Europe. The time-of-flight, angular difference, associated-particle coincidence and lead filtration technique have been compared, using the same Geiger-Müller counter. For 15 MeV neutrons, a consistent kU-value of 1.9% was obtained. For the lower neutron energies, lower kU values were observed with reasonable agreement. |
---|---|