Quantitative separation of estrogens, androgens and progestogens using reverse phase high pressure liquid chromatography
Dual UV wavelength scanning at 206 and 254 nm was used to develop a sensitive, separation and quantitation procedure for estrone, estradiol-17β, testosterone, 4-androstene-3,17,-dione, progesterone and 17-hydroxyprogesterone using reverse phase high pressure liquid chromatography. The difference in...
Gespeichert in:
Veröffentlicht in: | Steroids 1982-10, Vol.40 (4), p.381-388 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Dual UV wavelength scanning at 206 and 254 nm was used to develop a sensitive, separation and quantitation procedure for estrone, estradiol-17β, testosterone, 4-androstene-3,17,-dione, progesterone and 17-hydroxyprogesterone using reverse phase high pressure liquid chromatography. The difference in UV absorption of the estrogens from the androgens and progestogens allows for correction of the co-elution of testosterone with estradiol-17β and 4-androsten-3,17,-dione with estrone. Incorporation of an isocratic step-gradient provides improved separation while maintaining shortened elution times for the less polar steroids. |
---|---|
ISSN: | 0039-128X 1878-5867 |
DOI: | 10.1016/0039-128X(82)90015-0 |