Technique modification for radioactive implants when plastic tubes are used in fibrosed head and neck areas
Severe fibrosis around an area implanted with plastic tubes required a strong pulling tension force to place them in position. This pull produced a stretching of the tubes, and a consequent narrowing of the lumen, making afterloading with radioactive seeds impossible. Comments about this problem are...
Gespeichert in:
Veröffentlicht in: | Int. J. Radiat. Oncol., Biol. Phys.; (United States) Biol. Phys.; (United States), 1982-10, Vol.8 (10), p.1799-1801 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Severe fibrosis around an area implanted with plastic tubes required a strong pulling tension force to place them in position. This pull produced a stretching of the tubes, and a consequent narrowing of the lumen, making afterloading with radioactive seeds impossible. Comments about this problem are made. Regular techniques in this situation cannot be used, and a technique modification is proposed. |
---|---|
ISSN: | 0360-3016 1879-355X |
DOI: | 10.1016/0360-3016(82)90305-4 |