Simplified assay of amrinone in plasma by high-performance liquid chromatography
An isocratic high-performance liquid chromatography method for the assay of plasma amrinone is described. Plasma amrinone is extracted using protein precipitation with an internal standard, separated with a reverse phase column, and detected using ultraviolet absorption. Each run is completed within...
Gespeichert in:
Veröffentlicht in: | Therapeutic drug monitoring 1990-11, Vol.12 (6), p.570-573 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An isocratic high-performance liquid chromatography method for the assay of plasma amrinone is described. Plasma amrinone is extracted using protein precipitation with an internal standard, separated with a reverse phase column, and detected using ultraviolet absorption. Each run is completed within 10 min. The assay can detect amrinone concentrations between 0.5 and 10.0 micrograms/ml, within the accepted therapeutic range. The assay has a within-day coefficient of variation of less than 5% and a day-to-day coefficient of variation of less than 10% in the therapeutic range of amrinone. This technique is an accurate, simple, and rapid method for the determination of amrinone concentrations in plasma. |
---|---|
ISSN: | 0163-4356 1536-3694 |
DOI: | 10.1097/00007691-199011000-00010 |