Evaluation of a simple method for the measurement of cytidine deaminase in serum and comparison with a reference method
This study was designed to evaluate a cytidine deaminase (CD) assay modified to allow results to be achieved within one working day. Inter-batch variation for samples of mean (SD) CD activity, 10.2 (1.0) units, 17.5 (1.2) units and 31.7 (1.7) units were, 9.8%, 6.9% and 5.4% respectively ( n = 26). T...
Gespeichert in:
Veröffentlicht in: | Clinica chimica acta 1990-11, Vol.192 (1), p.55-59 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study was designed to evaluate a cytidine deaminase (CD) assay modified to allow results to be achieved within one working day. Inter-batch variation for samples of mean (SD) CD activity, 10.2 (1.0) units, 17.5 (1.2) units and 31.7 (1.7) units were, 9.8%, 6.9% and 5.4% respectively (
n = 26). The reference range (3.2–13.2 U) was similar in males and females and was independent of age. There was close correlation with a reference method (
r = 0.96). The mean difference between methods was 2.7 U and the limits of agreement were − 1.7 to 7.1 U. The results indicate that the short assay technique can produce results that are sufficiently accurate and precise to be clinically useful. |
---|---|
ISSN: | 0009-8981 1873-3492 |
DOI: | 10.1016/0009-8981(90)90271-S |