Model to describe the degree of twitch potentiation during neuromuscular monitoring

Neuromuscular block is estimated by comparing the evoked peak twitch with a control value measured in the absence of neuromuscular block. In practice, this control value is often difficult to determine because repeated motor nerve stimulation enhances the evoked mechanical response of the correspond...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 2004-03, Vol.92 (3), p.373-380
Hauptverfasser: Eleveld, D.J., Kopman, A.F., Proost, J.H., Wierda, J.M.K.H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Neuromuscular block is estimated by comparing the evoked peak twitch with a control value measured in the absence of neuromuscular block. In practice, this control value is often difficult to determine because repeated motor nerve stimulation enhances the evoked mechanical response of the corresponding muscle, resulting in an increased twitch response. This is known as twitch potentiation or the staircase phenomenon. It is probably the result of myosin light chain phosphorylation creating an increased twitch force for a given amount of Ca2+ released at each action potential. Modelling of potentiation may improve studies of neuromuscular blocking agents using mechanomyography or accelerometry. We used one- and two-exponential models to describe the degree of myosin light chain phosphorylation and associated twitch potentiation. These models were fitted to accelerographic twitch force measurements for various stimulation patterns and frequencies used in neuromuscular monitoring. Fitting a two-exponential model to twitch data for various stimulation rates and patterns provides better prediction than a one-exponential model. A one-exponential model performs poorly when the stimulation rate varies during measurement. We conclude that a two-exponential model can predict the degree of twitch potentiation for the stimulation patterns and frequencies tested more accurately than a one-exponential model. However, if only one stimulation frequency is used, a one-exponential model can provide good accuracy. We illustrate that such a potentiation model can improve the ability of pharmacodynamic-pharmacokinetic neuromuscular block models to predict twitch response in the presence of a neuromuscular blocking agent.
ISSN:0007-0912
1471-6771
DOI:10.1093/bja/aeh056