Positronium: Review of symmetry, conserved quantities and decay for the radiological physicist

The PET-CT scanner, which overlays a high spatial resolution CT image on a relatively low resolution physiologic PET image, has greatly increased the efficacy of the PET imaging modality. With the resulting increase in the presence of PET imaging in the radiologic and medical physics communities, me...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2004-01, Vol.31 (1), p.57-61
1. Verfasser: Harpen, Michael D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The PET-CT scanner, which overlays a high spatial resolution CT image on a relatively low resolution physiologic PET image, has greatly increased the efficacy of the PET imaging modality. With the resulting increase in the presence of PET imaging in the radiologic and medical physics communities, medical physicists can expect professional responsibilities in acceptance testing, quality assurance, radiation safety and resident teaching of this new modality. With this in mind we present a review of a subtle aspect of PET: the formation, symmetry properties and decay of positronium; the hydrogen like positron–electron complex which forms prior to annihilation.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.1630494