Transesophageal doppler ultrasonography : evidence for improved cardiac output monitoring
Cardiac output monitoring by transesophageal Doppler ultrasound has not gained wide clinical acceptance. A recently developed transesophageal Doppler device, Accucom 2, features technological advances aimed to reduce the error of this approach to the monitoring of cardiac output. To determine if Acc...
Gespeichert in:
Veröffentlicht in: | Anesthesia and analgesia 1990-12, Vol.71 (6), p.651-657 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Cardiac output monitoring by transesophageal Doppler ultrasound has not gained wide clinical acceptance. A recently developed transesophageal Doppler device, Accucom 2, features technological advances aimed to reduce the error of this approach to the monitoring of cardiac output. To determine if Accucom 2 enables more accurate cardiac output monitoring, a prospective investigation was undertaken. Analyses of 923 simultaneous measurements of transesophageal Doppler and thermodilution cardiac output in 47 anesthetized patients were compared. Results using Accucom 2 were highly correlated with the results obtained with thermodilution (R = 0.91) and provided significantly greater accuracy in comparison with a previously developed Doppler cardiac output monitor, Accucom 1. Technological advances in transesophageal Doppler ultrasound hold promise for clinically useful noninvasive monitoring of cardiac output. |
---|---|
ISSN: | 0003-2999 1526-7598 |
DOI: | 10.1213/00000539-199012000-00013 |