Method development for determining the iohexol in human serum by micellar electrokinetic capillary chromatography
Iohexol is widely used in clinical laboratories as a non-ionic radiographic contrast medium. Determination of its concentration in blood has a vital meaning in preventing its side effects caused by its retention in the system. A method for determining iohexol in serum by micellar electrokinetic capi...
Gespeichert in:
Veröffentlicht in: | Journal of pharmaceutical and biomedical analysis 2004-01, Vol.34 (1), p.153-158 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Iohexol is widely used in clinical laboratories as a non-ionic radiographic contrast medium. Determination of its concentration in blood has a vital meaning in preventing its side effects caused by its retention in the system. A method for determining iohexol in serum by micellar electrokinetic capillary chromatography (MECC) requiring no pretreatment is developed. Electrophoresis is performed for serum samples at 25
kV with a borate buffer (50
mM; pH 9.5) containing sodium dodecyl sulfate (50
mM) and detection is carried out at 245
nm. Migration time of iohexol is 7.4
min. Linearity (0–1000
mg/l) is good and detection limit is 0.5
mg/l (
S/
N=3). CV of intra-assay precision at a measurement concentration range of 6.2–200.1
mg/l is 1.38–4.68% and recovery rate is 96–102%. CV of inter-assay precision is 2.06–5.94% at a measurement concentration range of 10.3–155.4
mg/l. This method is characterized by determination through direct injection of serum samples of super micro-quantity into the capillary, which simplifies the determination procedure in a significant manner and improves the precision and accuracy of determination. |
---|---|
ISSN: | 0731-7085 1873-264X |
DOI: | 10.1016/j.japna.2003.08.019 |