A high pressure matrix-assisted laser desorption ion source for Fourier transform mass spectrometry designed to accommodate large targets with diverse surfaces
A new design for a high pressure MALDI-FTMS instrument is described and initial data are shown. The instrument incorporates a large, 10 cm × 10 cm, sample translation stage to accommodate and position the MALDI target. The new instrument allows coupling to a wide variety of surface techniques such a...
Gespeichert in:
Veröffentlicht in: | Journal of the American Society for Mass Spectrometry 2004, Vol.15 (1), p.128-132 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new design for a high pressure MALDI-FTMS instrument is described and initial data are shown. The instrument incorporates a large, 10 cm × 10 cm, sample translation stage to accommodate and position the MALDI target. The new instrument allows coupling to a wide variety of surface techniques such as gel electrophoresis or surface plasmon resonance. Coupling to thin layer chromatography is shown. Furthermore, a new nozzle design allows high pressure collisional cooling sufficient to stabilize gangliosides while minimizing the gas load on the system. |
---|---|