Leukocyte depletion filters: A comparison of efficiency
Four commercially available leukocyte depletion filters were evaluated using a flow cytometric technique to determine the efficiency of each type of filter. The Sepacell R‐500A® and the PALL RC50® were the most efficient in leukocyte depletion having a mean depletion percentage of 99.3% and 99.5%, r...
Gespeichert in:
Veröffentlicht in: | American journal of hematology 1990-11, Vol.35 (3), p.184-186 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Four commercially available leukocyte depletion filters were evaluated using a flow cytometric technique to determine the efficiency of each type of filter. The Sepacell R‐500A® and the PALL RC50® were the most efficient in leukocyte depletion having a mean depletion percentage of 99.3% and 99.5%, respectively. The PALL RC100® which is used for two units also had a 99.3% depletion with the first unit but with the second unit the depletion dropped to 94.2%. The Imugard IG‐500® had a 97.3% depletion. Red cell recovery ranged from 87.4% for the PALL RC50® to 92.2% for the Sepacell R‐500A®. |
---|---|
ISSN: | 0361-8609 1096-8652 |
DOI: | 10.1002/ajh.2830350308 |