The validity of continuous automated fluid monitoring during endometrial surgery: luxury or necessity?

Thirty‐four consecutive women undergoing endometrial laser ablation, as a treatment of menorrhagia, were recruited to assess the validity of fluid absorption monitoring by a new continuous automated system (AquaSens). The same group of women also had monitoring of fluid absorption carried out by our...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BJOG : an international journal of obstetrics and gynaecology 1998-07, Vol.105 (7), p.797-801
Hauptverfasser: Hawe, Jeremy A., Chien, Patrick F. W., Martin, Doreen, Graham Phillips, A., Garry, Ray
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Thirty‐four consecutive women undergoing endometrial laser ablation, as a treatment of menorrhagia, were recruited to assess the validity of fluid absorption monitoring by a new continuous automated system (AquaSens). The same group of women also had monitoring of fluid absorption carried out by our standard technique of weighing. The intra‐class correlation coefficient for the fluid deficit estimated by AquaSens compared to our standard technique of manually weighing the irrigation bags was 0.98 (95% CI 0.96–0.99). Aquasens therefore provides a valid and non‐invasive method of continuously monitoring fluid deficit amongst patients undergoing operative hysteroscopy procedures, thereby reducing the risk of unexpected fluid absorption and its potentially fatal sequelae.
ISSN:1470-0328
0306-5456
1471-0528
1365-215X
DOI:10.1111/j.1471-0528.1998.tb10213.x