Instrumented measurement of the posterolateral corner
A new device called the Lars Rotational Laxiometer (Lars Inc, Dijon, France) is introduced to aid in the diagnosis of posterolateral rotatory instability of the knee. This device assigns a quantitative value for tibial external rotation. Three examiners each evaluated a separate group of 30 differen...
Gespeichert in:
Veröffentlicht in: | Arthroscopy 1998-07, Vol.14 (5), p.489-494 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new device called the Lars Rotational Laxiometer (Lars Inc, Dijon, France) is introduced to aid in the diagnosis of posterolateral rotatory instability of the knee. This device assigns a quantitative value for tibial external rotation. Three examiners each evaluated a separate group of 30 different subjects (total 180 knees) to obtain side-to-side differences. The subjects had no history of injury, pain, or instability. An external rotation measurement was performed at 30 degrees and 90 degrees of knee flexion. At 90 degrees, the mean side-to- side difference was 4.4 degrees (range, 3.7 degrees to 5.1 degrees); at 30 degrees it was 5.5 degrees (range, 4.7 degrees to 6.3 degrees). There was no significant difference with gender or age. The purpose of this study is to establish baseline side-to-side values for the posterolateral complex in normal knees. Objective values are obtainable with the Laxiometer.
Arthroscopy 1998 Jul-Aug;14(5):489-94 |
---|---|
ISSN: | 0749-8063 1526-3231 |
DOI: | 10.1016/S0749-8063(98)70077-5 |