Quantification of Aortic Stenosis in Mechanically Ventilated Patients Using Multiplane Transesophageal Doppler Echocardiography
To evaluate the feasibility and accuracy of multiplane transesophageal Doppler echocardiographic assessment of the severity of aortic stenosis in mechanically ventilated patients using modified transgastral views of the left ventricular outflow tract and the aortic valve. A prospective study compari...
Gespeichert in:
Veröffentlicht in: | Chest 1998-07, Vol.114 (1), p.94-97 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the feasibility and accuracy of multiplane transesophageal Doppler echocardiographic assessment of the severity of aortic stenosis in mechanically ventilated patients using modified transgastral views of the left ventricular outflow tract and the aortic valve.
A prospective study comparing the results of transesophageal echocardiography (TEE) with transthoracic echocardiography (TTE) and cardiac catheterization.
A university hospital.
Twenty-eight American Society of Anesthesiologists class III and IV patients with aortic stenosis undergoing elective cardiac surgery for valve replacement.
Intubated and mechanically ventilated patients with aortic stenosis undergoing cardiac surgery for valve replacement were studied by multiplane transesophageal Doppler echocardiography to determine transvalvular pressure gradients (Bernoulli formula) and valve areas (continuity equation). These findings were compared with the respective preoperative data from TTE and cardiac catheterization.
In 25 of 28 patients (89%), adequate transgastral Doppler recordings of the aortic jet could be obtained. The TEE measurements correlated well with the respective data obtained by TTE (maximal pressure gradient: r=0.93, p |
---|---|
ISSN: | 0012-3692 1931-3543 |