Tonometry in adults and children: A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo
The purpose of the study was to determine the accuracy of applanation tonometry, pneumatonometry, and TonoPen tonometry in adults and children and the effect of age on tonometer error. The design was divided into four parts: part 1 was prospective and cross-sectional, and parts 2 through 4 were pros...
Gespeichert in:
Veröffentlicht in: | Ophthalmology (Rochester, Minn.) Minn.), 1998-07, Vol.105 (7), p.1173-1181 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of the study was to determine the accuracy of applanation tonometry, pneumatonometry, and TonoPen tonometry in adults and children and the effect of age on tonometer error.
The design was divided into four parts: part 1 was prospective and cross-sectional, and parts 2 through 4 were prospective, cross-sectional, and masked.
This study contained 72 patients representing 74 data points.
Tonometry with simultaneous manometry was performed.
Intraocular pressure (IOP) and the tonometric estimate of IOP were obtained.
The normal pediatric IOP follows the line Ta = 0.71 age(years) + 10 up to age 10. Applanation tonometry under anesthesia differs from pneumatonometry by an average of −8.6 mmHg and is age related by the equation Ta = Tpn + 2.6 log(age) − 10.3. The TonoPen was the most accurate instrument for enucleated eyes, and the pneumatonometer was the most accurate in anesthetized living eyes.
Applanation tonometry markedly underestimated IOP in young eyes. TonoPen tonometry performed well with enucleated eyes but was not adequately accurate for clinical use. The pneumatonometer performed the best clinically and the best overall. |
---|---|
ISSN: | 0161-6420 1549-4713 |
DOI: | 10.1016/S0161-6420(98)97016-6 |