Determination of alanine aminotransferase in human serum in an open-closed flow injection configuration
Open-closed flow injection systems allow the development of methods based on multidetection by using conventional detectors. The methods proposed here for the determination of alanine aminotransferase (ALT) utilize a photometric detector included in the closed circuit, which allows one to perform fi...
Gespeichert in:
Veröffentlicht in: | Journal of biotechnology 1990-04, Vol.14 (1), p.43-52 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Open-closed flow injection systems allow the development of methods based on multidetection by using conventional detectors. The methods proposed here for the determination of alanine aminotransferase (ALT) utilize a photometric detector included in the closed circuit, which allows one to perform fixed-time and reaction-rate measurements whose sensitivity can be increased by using the sum of the analytical signals. The methods, with determination ranges between 1 and 500 U l
−1 of ALT, feature sampling frequencies up to 60 h
−1. Their application to serum samples provided excellent results, with recoveries between 95 and 106% and averaging at 99.9%. |
---|---|
ISSN: | 0168-1656 1873-4863 |
DOI: | 10.1016/0168-1656(90)90016-5 |