Determination of alanine aminotransferase in human serum in an open-closed flow injection configuration

Open-closed flow injection systems allow the development of methods based on multidetection by using conventional detectors. The methods proposed here for the determination of alanine aminotransferase (ALT) utilize a photometric detector included in the closed circuit, which allows one to perform fi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biotechnology 1990-04, Vol.14 (1), p.43-52
Hauptverfasser: Fernández-Romero, J.M., Luque de Castro, M.D., Valcárcel, M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Open-closed flow injection systems allow the development of methods based on multidetection by using conventional detectors. The methods proposed here for the determination of alanine aminotransferase (ALT) utilize a photometric detector included in the closed circuit, which allows one to perform fixed-time and reaction-rate measurements whose sensitivity can be increased by using the sum of the analytical signals. The methods, with determination ranges between 1 and 500 U l −1 of ALT, feature sampling frequencies up to 60 h −1. Their application to serum samples provided excellent results, with recoveries between 95 and 106% and averaging at 99.9%.
ISSN:0168-1656
1873-4863
DOI:10.1016/0168-1656(90)90016-5