A comparison of intubation success for paediatric transport team paramedics using lighted vs regular tracheal tube stylets

We conducted a prospective randomized study of success rate and time to intubation using Trachlight and Surch‐Lite lighted stylets versus a regular tracheal tube stylet, in a training setting. Participants, 18 paediatric transport paramedics, performed two intubations with each of the three devices,...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pediatric anesthesia 1998-05, Vol.8 (3), p.215-220
Hauptverfasser: MACNAB, ANDREW J., MACPHAIL, IAIN, MACNAB, MAGNUS K., NOBLE, ROBERT, O'FLAHERTY, DEBRA
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We conducted a prospective randomized study of success rate and time to intubation using Trachlight and Surch‐Lite lighted stylets versus a regular tracheal tube stylet, in a training setting. Participants, 18 paediatric transport paramedics, performed two intubations with each of the three devices, using an airway management trainer. There was no significant difference in mean time for intubation between the three devices. The times for external confirmation of correct tube placement were comparable using the two lighted stylets. External confirmation of the tube placement using the lighted stylets was quicker than laryngoscopic visualization. In darkness, with a nonfunctioning laryngoscope, intubations were successfully performed 100% of the time with the lighted stylet, but only 11% of the time with the regular stylet. All paramedics felt that a lighted stylet would be a useful airway management adjunct for the transport environment for complicated intubations or for use in very high or low levels of ambient light.
ISSN:1155-5645
1460-9592
DOI:10.1046/j.1460-9592.1998.00742.x