Advantages of Subharmonic Over Second Harmonic Backscatter for Contrast-To-Tissue Echo Enhancement

It is shown experimentally that backscatter from two ultrasonic contrast agents suspended in water or saline contains subharmonics of the incident frequency that are stronger than those backscattered at the same incident pressure from chicken breast. It is also shown that the ratio of subharmonic ba...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultrasound in medicine & biology 1998-03, Vol.24 (3), p.395-399
Hauptverfasser: Shankar, P.M, Krishna, P.Dala, Newhouse, V.L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:It is shown experimentally that backscatter from two ultrasonic contrast agents suspended in water or saline contains subharmonics of the incident frequency that are stronger than those backscattered at the same incident pressure from chicken breast. It is also shown that the ratio of subharmonic backscattered from contrast to that backscattered from tissue, is stronger than the ratio of backscattered second harmonic. In consequence, blood that contains contrast should be more easily detectable with respect to tissue if the subharmonic, rather than the second harmonic, is used for imaging.