Comparison of oscillation with three other methods for measuring nasal airways resistance
In this study we have compared the sensitivity and reproducibility of nasal airways resistance measurements made using an oscillometer, with those made by passive anterior, active anterior and active posterior rhinomanometry. Nasal airways resistance values were compared in 12 patients with rhinitis...
Gespeichert in:
Veröffentlicht in: | Respiratory medicine 1990-03, Vol.84 (2), p.101-106 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this study we have compared the sensitivity and reproducibility of nasal airways resistance measurements made using an oscillometer, with those made by passive anterior, active anterior and active posterior rhinomanometry. Nasal airways resistance values were compared in 12 patients with rhinitis and 15 normal subjects, of whom ten had additional measurements after a vasoconstrictor spray, oxymetazoline. The coefficients of variation of 6–8 technically satisfactory measurements were 9–19%. The decongestant effect of oxymetazoline was detected by all methods, with no decrease in reproducibility. Post vasoconstrictor nasal airways resistance fell by 28% (passive anterior), 35% (active anterior), 36% (active posterior) and 58% (oscillometry). In conclusion, the oscillation method for deriving nasal airways resistance is a useful, new, simple and noninvasive way of assessing nasal airways patency. Results compare favourably with other, more established techniques. |
---|---|
ISSN: | 0954-6111 1532-3064 |
DOI: | 10.1016/S0954-6111(08)80010-8 |