Optimizing Bone Screw Pullout Force
Stability of internal fixation by plate and screw depends on the interface of bone and screw threads. Bone-screw thread design was tested in a synthetic bone model to determine maximum bone-screw pullout force. The model was confirmed to provide reproducible results in a clinically relevant range. C...
Gespeichert in:
Veröffentlicht in: | Journal of orthopaedic trauma 1990, Vol.4 (2), p.169-174 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Stability of internal fixation by plate and screw depends on the interface of bone and screw threads. Bone-screw thread design was tested in a synthetic bone model to determine maximum bone-screw pullout force. The model was confirmed to provide reproducible results in a clinically relevant range. Consistent material properties were reflected in the relatively small standard deviations in pullout force, which were less than 10% of mean values. As expected, major diameter was an important determinant of pullout force in a roughly linear manner. Pitch was important with a finer thread giving greater purchase. Minor diameter and the ratio of major to minor diameter had a small but significant effect on pullout force. This study was significant for establishing a good bone-screw testing model and suggesting important thread parameters in selecting or manufacturing a bone screw to optimize its holding power. |
---|---|
ISSN: | 0890-5339 1531-2291 |
DOI: | 10.1097/00005131-199004020-00012 |