Comparative Accuracy of Three Automated Techniques in the Noninvasive Estimation of Central Blood Pressure in Men

Automated devices have regularly replaced manual sphygmomanometry for the determination of blood pressure not only in homes and clinics, but also in emergency and critical care settings. Few studies exist that correctly assess the accuracy of these devices, and even fewer that specifically compare c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The American journal of cardiology 1998-04, Vol.81 (8), p.1004-1012
Hauptverfasser: Lehmann, Kenneth G, Gelman, Julie A, Weber, Michael A, Lafrades, Avelino
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Automated devices have regularly replaced manual sphygmomanometry for the determination of blood pressure not only in homes and clinics, but also in emergency and critical care settings. Few studies exist that correctly assess the accuracy of these devices, and even fewer that specifically compare commercially available units that rely on different physiologic events for measurement. Six hundred pressure measurements were obtained from 120 subjects using 1 of 3 randomly selected blood pressure monitors. In addition, central arterial pressure measurements were obtained simultaneously and directly from the ascending aorta of each subject. Overall, these devices tended to overestimate diastolic (+2.5 mm Hg, p