Comparison of four commercially available 81Rb- 81mKr generators: Quality testing using a breathing lung phantom

Four commercially available 81Rb- 81mKr generators, including their elution systems, were compared. Considerable differences in useful output between the various generators were found. When a standardized elution system was used, the differences became much less. Differences in the tubing and inaccu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of radiation applications and instrumentation. Part B, Nuclear medicine and biology Nuclear medicine and biology, 1990, Vol.17 (3), p.269-272
Hauptverfasser: Camps, Johannes A.J., Blokland, Jacobus A.K., Blok, Dik, Pauwels, Ernest K.J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Four commercially available 81Rb- 81mKr generators, including their elution systems, were compared. Considerable differences in useful output between the various generators were found. When a standardized elution system was used, the differences became much less. Differences in the tubing and inaccuracy of the flow meters were found to be the major causes of the original differences. The user should be aware of a loss of output that may occur when the generator is eluted with dry air.
ISSN:0883-2897
DOI:10.1016/0883-2897(90)90051-2