Development and validation of an automated enzyme assay for paracetamol (acetaminophen)
A rapid, enzymatic assay for serum or plasma paracetamol has been developed with the potential for adaptation to a wide range of clinical analysers. The method involves the action of an amidase enzyme to produce 4-aminophenol from paracetamol, which in turn reacts with 8-hydroxyquinoline in the pres...
Gespeichert in:
Veröffentlicht in: | Clinica chimica acta 1990-02, Vol.187 (2), p.95-104 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A rapid, enzymatic assay for serum or plasma paracetamol has been developed with the potential for adaptation to a wide range of clinical analysers. The method involves the action of an amidase enzyme to produce 4-aminophenol from paracetamol, which in turn reacts with 8-hydroxyquinoline in the presence of manganese ions to form a blue dye. Two stable reagents are used and excellent precision is achieved over the drug concentration range 0–2.5 mmol/l. The method, which is complete within 6 min, has been validated using a Monarch
™ centrifugal analyser and shows no significant interference from endogenous serum compounds, drugs or paracetamol metabolites. |
---|---|
ISSN: | 0009-8981 1873-3492 |
DOI: | 10.1016/0009-8981(90)90335-P |