Incidence of lightspeed separation and the potential for bypassing

It is important for endodontic instruments to have a low fracture rate. If a fracture does occur, it would be desirable to have the ability to bypass the broken segment and complete the root canal treatment. One hundred sixty-two root canals in 52 maxillary and mandibular first molars were cleaned a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of endodontics 1997-09, Vol.23 (9), p.586-587
Hauptverfasser: Ramirez-Salomon, Marco, Soler-Bientz, Rolando, de la Garza-González, Rafael, Palacios-Garza, Claudia M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:It is important for endodontic instruments to have a low fracture rate. If a fracture does occur, it would be desirable to have the ability to bypass the broken segment and complete the root canal treatment. One hundred sixty-two root canals in 52 maxillary and mandibular first molars were cleaned and shaped with Lightspeed instruments by three endodontists in their private practices. The canals were instrumented using the technique recommended by the manufacturer. All canals were instrumented to at least a size 45 at the working length. Six instruments separated during treatment. All six had been used more times than recommended by the manufacturer. Five of the six were easily bypassed and treatment completed.
ISSN:0099-2399
1878-3554
DOI:10.1016/S0099-2399(06)81128-3